V. Narasimhadhan, A. and Sharma, Aman and Mistry, Dipen (2013) Image Reconstruction from Fan-Beam Projections without Back-Projection Weight in a 2-D Dynamic CT: Compensation of Time-Dependent Rotational, Uniform Scaling and Translational Deformations. Open Journal of Medical Imaging, 03 (04). pp. 136-143. ISSN 2164-2788
OJMI_2013120914525767.pdf - Published Version
Download (717kB)
Abstract
In a dynamic CT, the acquired projections are corrupted due to strong dynamic nature of the object, for example: lungs, heart etc. In this paper, we present fan-beam reconstruction algorithm without position-dependent backprojection weight which compensates for the time-dependent translational, uniform scaling and rotational deformations occurring in the object of interest during the data acquisition process. We shall also compare the computational cost of the proposed reconstruction algorithm with the existing one which has position-dependent weight. To accomplish the objective listed above, we first formulate admissibility conditions on deformations that is required to exactly reconstruct the object from acquired sequential deformed projections and then derive the reconstruction algorithm to compensate the above listed deformations satisfying the admissibility conditions. For this, 2-D time-dependent deformation model is incorporated in the fan-beam FBP reconstruction algorithm with no backprojection weight, assuming the motion parameters being known. Finally the proposed reconstruction algorithm is evaluated with the motion corrupted projection data simulated on the computer.
Item Type: | Article |
---|---|
Subjects: | Middle East Library > Medical Science |
Depositing User: | Unnamed user with email support@middle-eastlibrary.com |
Date Deposited: | 29 Mar 2023 07:04 |
Last Modified: | 25 May 2024 09:19 |
URI: | http://editor.openaccessbook.com/id/eprint/358 |