Hu, Minxia and Zhao, Xinming and Song, Junfeng and Zhou, Chunwu (2013) Reduction of the Radiation Dose by Decreasing the Tube Current without Degradation of Low-Contrast Detectability on Abdominal Multi-Detector Row CT: A Phantom-Based Study. Open Journal of Medical Imaging, 03 (04). pp. 110-115. ISSN 2164-2788
OJMI_2013120616005432.pdf - Published Version
Download (251kB)
Abstract
As use of the lowest acceptable radiation dose during routine diagnostic imaging is important, we determined the optimal tube current without degradation of low-contrast detectability on abdominal multi-detector row CT (MDCT). CT scanning was performed with a Catphan® 500 phantom. The optimal tube current was 300 mA on 64-MDCT and 160 mA on 8-MDCT, with a fixed voltage of 120 kV. Reduction of the radiation dose in abdominal CT scanning by lowering the tube current proved to be feasible.